Extech Instruments EX505 User Manual Page 3

  • Download
  • Add to my manuals
  • Print
  • Page
    / 14
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 2
EX505 V6.1 11/07
3
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function Maximum Input
V DC or V AC 1000VDC/AC rms
mA AC/DC 400mA AC/DC
A AC/DC 10A AC/DC (20A for 30 sec)
Frequency, Resistance, Capacitance,
Diode Test, Continuity, Temperature
600VDC/AC rms
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above
earth ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in
the current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when
making resistance or diode tests.
6. ALWAYS turn off the power and disconnect the test leads before opening the covers to
replace the fuse or batteries.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in
place and fastened securely.
8. If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Page view 2
1 2 3 4 5 6 7 8 ... 13 14

Comments to this Manuals

No comments