Page view 6
Lo Limit
Pin 3 (+)
Pin 4 (-)
Figure 4
Hi Limit
Pin 1 (+)
Pin 2 (-)
Figure 3
Pin 5
Pin 1
Figure 1
Pin 2
Pin 4
Pin 6
Pin 3
Pin 5 (+)
Pin 6 (-)
Figure 2
5to30V
Setup Mode
The operating functions for the setup mode are printed between the bracket “[ ]” signs on
each button. Press the SET [ ] button in normal mode to switch to the setup mode. The
indicator “SET” will be shown on the left side of the display. Press SET [ ] again to return to
normal mode.
Hi/Lo Limit Setup
In Setup Mode, press the [Limits] button to enter the Hi/Lo setup function. The “LIMIT”,
“HI”, and main display will blink on the LCD and the previous settings will be displayed.
Press the number buttons (printed in white) on the overlay to set Hi and Lo limit values.
The button “-“ (same as the ESC overlay) can be used to set a negative value. Setting is
from left to right digit. Press the ENTER button to confirm each setting.
Mini DIN Hi/Lo Alarm Output
The alarm output jack (refer to Figure 1), located on the top of the meter, can be used to
power external warning or switching devices when programmed Hi/Lo alarm limits are
reached. To enable this function, an external power source (5 to 30VDC, 5mA max) must
be used to supply power to the connected device. Connect the power source (5-30VDC)
between pin 5 (+) and pin 6 (-) of the connector (refer to Figure 2). For the Hi limit output,
connect a device between pin 1 (+) and pin 2 (-) (refer to Figure 3). For the Lo limit output,
connect a device between pin 3 (+) and pin 4 (-) (refer to Figure 4).