Enclosure Double molded
Diode Test Test current of 0.9mA maximum, open circuit voltage 2.8V DC typical
Continuity Threshold 20 to 50, test current <1.5mA
Input Impedance 10M VDC/VAC
AC Response True rms
ACV Bandwidth 40Hz to 1000Hz
Display 2,000 count backlit liquid crystal display
Over-range indicationOL” is displayed
Auto Power Off 15 minutes (approximately)
Polarity Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Battery One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 200mA 250V ceramic fast blow
A range; 10A 600V ceramic fast blow
Operating Temperature -10ºC to 40ºC (14ºF to 122ºF)
Storage Temperature -10
o
C to 60
o
C (14
o
F to 140
o
F)
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF)
Storage Humidity <80%
Operating Altitude 2000 meters (7000 ft) maximum
Weight 260g (9.17oz)
Size 147 x 76 x 42mm (5.8 x 2.9 x 1.6”)
Safety This meter is intended for origin of installation use and protected, against
the users, by double insulation per IEC/EN 61010-1:2001 and IEC/EN
61010-031:2002 to Category III 600V; Pollution Degree 2.
Approvals CE
Copyright © 2012 Extech Instruments Corporation (a FLIR company)