Extech Instruments 382357 Specifications Page 4

  • Download
  • Add to my manuals
  • Print
  • Page
    / 13
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 3
382357‐EU‐ENV2.42/13
4
Operation
Note:Ensurethatthejawsarefullyclosedbeforetesting.
Note:DONOTclampontoanyconductororopentheclampjawsduringthestart‐upself‐
calibration(self‐calisidentifiedbytheLCDcountingdownfromCAL7toCAL1).
GroundResistanceMeasurements
1. Opentheclampjawsandcheckthatallsurfacesare
cleanandfreeofdust,dirt,oranyforeign
substances.Cleanifnecessary.
2. Forcefullysnaptheclampjawsopenandclosed
severaltimes.
3. Turnthemeteronbymovingtherotaryswitchto
theohms
position.DONOTinterruptthestartup
procedurebyopeningtheclampjawsorclamping
aroundaconductor.Duringself‐calibration,the
meterwilldisplayCAL5,CAL4,CAL3,CAL2and
CAL1.
4. Oncethestart‐upcalibrationprocedureiscompleted,themeterwillsoundabeep.Clamp
ontoanelectrodeorgroundrodandreadthevalueofground resistanceonthedisplay.
Ground/LeakageCurrentMeasurement
1. Turnthemeteronbymovingtherotaryswitchtothe
mAorAposition.
2. Clampontotheelectrodeorconductor
3. Readthevalueofleakagecurrentdisplayedonthe
LCD.
HOLD
PresstheHOLDbuttontofreezethereadinginthedisplay.Hwillappearinthedisplay.Press
HOLDagaintoexitthefunction.
Page view 3
1 2 3 4 5 6 7 8 9 ... 12 13

Comments to this Manuals

No comments